| Electrical Specifications | 
        
        
            | DC Voltage Ranges | 
            400 mV, 4 V, 40 V, 400 V, 1000 V | 
        
        
            | DC Voltage Accuracy | 
            all ranges: ± (0.7 % rdg + 1 dgt) | 
        
        
            | DC Voltage Resolution | 
            100 µV | 
        
        
            | DC Voltage Input Impedance | 
            10 Mω | 
        
        
            | AC Voltage (45 Hz to 500 Hz) Ranges | 
            400 mV, 4 V, 40 V, 400 V, 750 V | 
        
        
            | AC Voltage (45 Hz to 500 Hz) Accuracy | 
            4 V Range, 200 Hz to 500 Hz: ± (2.0 % rdg + 4 dgts) | 
        
        
            | AC Voltage (45 Hz to 500 Hz) Accuracy | 
            All other ranges: ± (1.5 % rdg + 4 dgts) | 
        
        
            | AC Voltage (45 Hz to 500 Hz) Resolution | 
            100 µV | 
        
        
            | AC Voltage (45 Hz to 500 Hz) Input Impedance | 
            10 Mω | 
        
        
            | DC Current Ranges | 
            400 µA, 4 mA, 40 mA, 300 mA, 10 A | 
        
        
            | DC Current Accuracy | 
            400 µA to 300 mA ranges: ± (1.0 % rdg + 1 dgt) | 
        
        
            | DC Current Accuracy | 
            10 A range: ± (2.0 % rdg + 3 dgts) | 
        
        
            | DC Current Resolution | 
            0.1 µA | 
        
        
            | DC Current Voltage Burden | 
            400 µA range: 1 mV/µA, 4 mA range: 100 mV/1 mA, 40 mA range: 12 mV/1 mA, | 
        
        
            | DC Current Voltage Burden | 
            300 mA range: 4 mV/1 mA, 10 A range: 100 mV/1 A | 
        
        
            | AC Current (45 Hz to 500 Hz) Ranges | 
            400 µA, 4 mA, 40 mA, 300 mA, 10 A | 
        
        
            | AC Current (45 Hz to 500 Hz) Accuracy | 
            400 µA to 300 mA ranges: ± (1.5 % rdg + 4 dgts) | 
        
        
            | AC Current (45 Hz to 500 Hz) Accuracy | 
            10 A range: ± (2.5 % rdg + 4 dgts) AC Current (45 Hz to 500 Hz) Resolution: 0.1 µA | 
        
        
            | AC Current (45 Hz to 500 Hz) Voltage Burden | 
            See DC Current | 
        
        
            | Resistance Ranges | 
            400 ω 40 kω 4 Mω | 
        
        
            | Resistance Accuracy | 
            400 ωand 40 kω ranges: ± (1.0 % rdg + 4 dgts) | 
        
        
            | Resistance Accuracy | 
            4 Mω range: ± (1.2 % rdg + 4 dgts) | 
        
        
            | Resistance Resolution | 
            100 mω | 
        
        
            | Resistance Open Circuit Voltage | 
            400 ω range: 3.0 V dc, All other ranges: 0.5 V dc typical | 
        
        
            | Frequency (autoranging) | 
        
        
            | Ranges | 
            4 kHz, 40 kHz, 400 kHz, 4 MHz, 40 MHz | 
        
        
            | Accuracy | 
            all ranges: ± (0.1 % rdg + 3 dgts) | 
        
        
            | Resolution | 
            1 Hz | 
        
        
            | Sensitivity | 
            10 Hz to 4 MHz: > 1.5 V ac rms 
            4 MHz to 40 MHz: > 2 V ac rms, < 5 V ac rms | 
        
        
            | Min Pulse Width | 
            > 25 ns | 
        
        
            | Duty Cycle Limits | 
            > 30 % and < 70 % | 
        
        
            | Continuity | 
        
        
            | Audible Indication | 
            < 35 ω | 
        
        
            | Response Time | 
            100 ms | 
        
        
            | Diode Test | 
        
        
            | Test Current | 
            1.0 mA (approximate) | 
        
        
            | Accuracy | 
            ± (1.5 % rdg + 3 dgts) | 
        
        
            | Resolution | 
            1 mV | 
        
        
            | Open Circuit Volts | 
            3.0 V dc typical | 
        
        
            | Peak Hold | 
        
        
            | AC Volts Accuracy | 
            40 V to 750 V: ± (3.0 % rdg + 60 dgts) | 
        
        
            | AC Current Accuracy | 
            ± (3.0 % rdg + 60 dgts) | 
        
        
            | Response Time | 
            > 1 ms | 
        
        
            | Overload Protection | 
        
        
            | AC/DC Voltage | 
            400 mV range: 1000 V dc or 750 V ac rms 15 sec 
            All other ranges: 1000 V dc or 750 V ac rms | 
        
        
            | AC/DC Current | 
            uA/mA input: 0.315 A /1000 V fast-blow ceramic (6.3 x 32 mm) 
            10 A input: 10 A /1000 V fast-blow ceramic (10 x 38 mm) 
            (10 A for 4 minutes maximum followed by a 12 minute cooling period) | 
        
        
            | Resistance | 
            1000 V dc or 750 V ac rms | 
        
        
            | Continuity | 
            1000 V dc or 750 V ac rms | 
        
        
            | Diode Test | 
            1000 V dc or 750 V ac rms | 
        
        
            | Capacitance | 
            0.315 A /1000 V fast-blow ceramic (6.3 x 32 mm) | 
        
        
            | Temperature | 
            1000 V dc or 750 V ac rms | 
        
        
            | Frequency | 