Fluke Calibration 5615-9BND-G
Catalog #: 4682907
Secondary PRT Probe, 100 OHM, 9 INCH 90 Deg. Bend
- Overview
- Specifications
- Summary
Product Highlights
- 5615-9BND-G Probe, Secondary PRT
- 100 OHM, 9IN 90 DEG BEND
- Affordable wide-range accuracy
- Calibrated accuracy ± 0.012 °C at 0 °C
- Reference-grade platinum sensing element
- NVLAP-accredited calibration included, lab code 200706-0
Each Unit Includes
- Coefficient Sheet
- User Manual CD
- Certification Sheets
Specifications |
Temperature range |
5615-9: –200 °C to 420 °C |
Nominal resistance at 0 °C |
100 Ω ± 0.10 Ω |
Temperature coefficient |
0.0039250 Ω/Ω/°C |
Accuracy[1] |
± 0.024 °C at –200 °C
± 0.012 °C at 0 °C
± 0.035 °C at 420 °CC |
Short-term repeatability[2] |
± 0.009 °C at 0.010 °C |
Drift[3] |
± 0.007 °C at 0.010 °C |
Sensor length |
28 mm (1.1 in) |
Sensor location |
6.9 mm ± 3.3 mm from tip (0.27 in ± 0.13 in) |
Sheath dimensions |
5615-9: 229 mm x 4.76 mm (9.0 in x 0.188 in) |
Sheath diameter tolerance |
± 0.127 mm (± 0.005 in) |
Sheath material |
Inconel™ 600 |
Minimum insulation resistance |
1000 M? at 23 °C |
Transition junction temperature range[4] |
–50 °C to 200 °C
|
Transition junction dimensions |
71 mm x 13 mm dia (2.8 in x 0.5 in) |
Maximum immersion length
|
5615-6: 102 mm (4 in)
5615-9: 178 mm (7 in)
5615-12: 254 mm (10 in) |
Response time[5] |
9 seconds typical |
Self heating (in 0 °C bath) |
50 mW/°C |
Lead-wire cable type |
PTFE insulated with PTFE jacket, 22 AWG |
Lead-wire length |
183 cm (72 in) |
Lead-wire temperature range |
–50 °C to 200 °C |
Calibration
|
NVLAP-accredited calibration included, lab code 200706-0. Please see calibration uncertainty table and its explanation of changeable uncertainties.
|
[1] Includes calibration and 100 hr drift (k=2)
[2] Three thermal cycles from min to max temp, includes hysteresis, 95 % confidence (k=2)
[3] After 100 hrs at max temp, 95 % confidence (k=2)
[4] Temperatures outside this range will cause irreparable damage. For best performance, transition junction should not be too hot to touch.
[5] Per ASTM E 644
|
NVLAP† Calibration Uncertainty |
Temperature |
Expanded Uncertainty (k=2) |
–196 °C |
0.024 °C |
–38 °C |
0.011 °C |
0 °C |
0.010 °C |
200 °C |
0.018 °C |
420 °C‡ |
0.029 °C |
Note: Calibration uncertainties depend on the uncertainties of the lab performing the calibration. Subsequent calibrations of this probe performed with different processes, at different facilities, or with changed uncertainty statements may state different uncertainties. †Lab code 200706-0 ‡5615-6 excluded
|
|