Fluke A3001 FC KIT
Catalog #: 4465665
FC Wireless Essential Kit With A3001
- Overview
- Specifications
- Summary
- Accessories
Product Highlights
- Save money and time by taking multiple measurements simultaneously
- Isolate interruptions or record signal fluctuations without even being there using the log function on the modules
- Work with fewer interruptions by recording multiple sessions before downloading
- Capture readings at different intervals, with user-set recording intervals from one second to one hour
- Choose the module(s) that work best for your applications—voltage, current, and temperature.
- View up to 10 live readings at once on the Fluke Connect™ mobile app
- View up to six live readings at once on a PC to get a complete understanding of the situation and download logging sessions (PC Adapter sold separately)
- Add the 3000 FC Digital Multimeter or Ti200/300/400 Infrared Camera to monitor modules on the move
- No more run-around. See multiple measurements in one place at the same time
- No more writing data down. Capture single or multiple logging sessions using the recording capability
- Avoid awkward or uncomfortable measurement positions. Set up the measurement once, step back and read it as many times as you want, when you want.
Fluke A3001 FC iFlex AC Current Module
- Measure up to 2500 A ac
- Use as a standalone meter or as part of the system
- Logging function for recording and saving up to
- 65,000 readings
- Inrush function
Fluke Connect Features:
- Fluke Connect allows you to wirelessly transfer data from your existing Fluke tools to your smart phone
- Fits over the IR port of your existing Fluke tools
- Upgradeable by USB
- Immediately enables you to graph, save, and share the tool’s readings with your team from your smart phone
Supported smart phones:
- Android™ 4.3 & up
- iPhone® 4s & later
Each Unit Includes
- (1) 3000 FC Wireless Digital Multimeter
- (1) Test lead set
- (1) Alligator clip set
- (1) A3001 FC Wireless iFlex AC Current Module
- (1) iFlex i2500-10 Flexible Current Probe
- (1) Magnetic Hanging strap
Fluke 3000 FC Series Wireless Multimeter |
Detailed specifications (Not compatible with Fluke CNX test tools)
For all specifications: accuracy is specified for one year after calibration, at operating temperatures of 18 °C to 28 °C, with relative humidity at 0 % to 90 %. Accuracy specifications take the form of ± ([ % of Reading ] + [ Number of least significant digits ]).
|
AC voltage |
Range 1 |
Resolution |
Accuracy2,3,4 |
45 Hz to 500 Hz |
500 Hz to 1 kHz |
600.0 mV |
0.1 mV |
1.0 % + 3 |
2.0 % + 3 |
6000 V |
0.001 mV |
1.0 % + 3 |
2.0 % + 3 |
60.00 V |
0.01 mV |
1.0 % + 3 |
2.0 % + 3 |
600.0 V |
0.1 mV |
1.0 % + 3 |
2.0 % + 3 |
1000 V |
1 mV |
1.0 % + 3 |
2.0 % + 3 |
1 All ac voltage ranges are specified from 1 % of range to 100 % of range.
2 Crest factor of = 3 at full scale up to 500 V, decreasing linearly to crest factor < 1.5 at 1000 V.
3 For non-sinusoidial waveforms, add ± (2 % of reading + 2 % full scale) typical, for crest factor up to 3.
4 Do not exceed 10 7 V-Hz. |
DC voltage, continuity, resistance, diode test and capacitance |
Function |
Range |
Resolution |
Accuracy |
mV |
600.0 mV |
0.1 mV |
0.09 % + 2 |
V |
6.000 V |
0.001 V |
0.09 % + 2 |
60.00 V |
0.01 V |
600.0 V |
0.1 V |
1000 V |
1 V |
0.15 % + 2 |
Ω |
600.0 Ω |
1 Ω |
Meter beeps at < 25 Ω, beeper detects opens or shorts of 250 us or longer. |
Diode test |
2.000 V |
0.001 V |
1 % + 2 |
µF |
1000 nF |
1 nF |
1.2 % + 2 |
10.00 µF |
0.01 µF |
100.0 µF |
0.1 µF |
9999 µF 1 |
1 µF |
10 % typical |
1 In the 9999 µF range for measurements to 1000 µF, the measurement accuracy is 1.2 % + 2. |
AC and DC current |
Function |
Range 1 |
Resolution |
Accuracy |
mA AC (45 Hz to 1 kHZ) |
60.00 mA |
0.01 mA |
1.5 % + 3 |
400.0 mA 3 |
0.1 mA |
mA DC 2 |
60.00 mA |
0.01 mA |
0.5 % + 3 |
400.0 mA 3 |
0.1 mA |
1All ac current ranges are specified from 5 % of range to 100 % of range.
2Input burden voltage (typical): 400 mA input 2 mV/m
3400.0 mA accuracy specified up to 600 mA overload. |
Frequency |
Range |
Resolution |
Accuracy 1 |
99.99 Hz |
0.01Hz |
0.1 % + 1 |
999.9 Hz |
0.1 Hz |
9.999 kHz |
0.001 kHz |
99.99 kHz |
0.01 kHz |
General specifications (Not compatible with Fluke CNX test tools) |
Maximum voltage between any terminal and earth ground |
1000 V dc or ac rms |
Ω fuse protection from A inputs |
0.44 A (44/100 A, 440 mA), 1000 V FAST Fuse, Fluke specified part only |
Display (LCD) |
Update rate: 4/sec
Votls, amps, ohms: 6000 counts
Frequency: 10,000 counts
Capacitance: 1,000 counts |
Battery type |
Three AA Alkaline batteries, NEDA 15A IEC LR6 |
Battery life |
250 hours minimum |
RF communications |
2.4 GHZ ISM Band |
RF communication range |
Open air, unobstructed: Up to 20m
Obstructed, sheetrock wall: Up to 6.5m
Obstructed, concrete wall, or steel electrical enclosure: Up to 3.5m |
Temperature |
Operating: -10 °C to 50 °C
Storage: -40 °C to 60 °C |
Temperature coefficient |
0.1 X (specified accuracy) /°C (<18 °C or >28 °C) |
Relative humidity |
0 % to 90 % (0 °C to 35 °C), 0 % to 75 % (35 °C to 40 °C), 0 % to 45 % (40 °C to 50 °C) |
Altitude |
Operating: 2,000 m
Storage: 12,000 m |
Electromagnetic compatibilty EMI, RFI, EMC, RF |
EN 61326-1:2006, EN 61326-2-2:2006. ETSI EN 300 328 V1.7.1:2006, ETSI EN 300 489 V1.8.1:2008, FCC Part 15 Subpart C Sections 15.207, 15.209, 15.249, FCCID : FCC: T68-FDMMBLE IC: 6627A-FDMMBLE |
Safety compliance |
ANSI/ISA 61010-1 / (82.02.01): 3 rd edition
CAN/CSA-C22.2 No 61010-1-12: 3 rd edition
UL 61010-1: 3 rd edition
IEC/EN 61010-1:2010 |
Certifications |
CSA, FCC, CE |
Ingress Protection (IP) rating |
IP54 |
Pollution Degree |
2 |
Size (H x W x L) |
4.75 cm x 9.3 cm x 20.7 cm (1.87 in x 3.68 in x 8.14 in) |
Weight |
340 g (12 oz) |
Type |
Digital |
Style |
Hand-Held |
Measures AC V |
Yes |
Measures DC V |
No |
Measures AC A |
Yes |
Measures DC A |
No |
True RMS |
Yes |
Maximum Voltage AC |
6000 V |
Minimum Voltage AC |
1 mV |
Resolution AC Volts |
600 mV |
Maximum Current AC |
60 mA (0.06 A) |
Minimum Current AC |
0.1 mA |
Resolution AC Current |
400 mA |
Maximum Current DC |
60 A |
Minimum Current DC |
0.1 mA |
Resolution DC Current |
400 mA |
Basic Accuracy V DC |
3 |
# Digit Display |
3.5 |
Backlight |
Yes |
Manual/Autoranging |
Both |
Dual Display |
No |
Analog Bargraph on Display |
No |
Decibels |
No |
Resistance |
Yes |
Frequency |
Yes |
Continuity |
Yes |
Capacitance |
Yes |
Temperature |
Yes |
Diode Testing |
Yes |
CAT Rating |
IV |
CAT Voltage |
1,000 V, 600 V |
Peak Hold |
No |
Overload Protector |
No |
Kelvin 4-Wire Measurement Capability |
No |
Intrinsically Safe |
No |
Wireless Display |
Yes |
Data Hold |
No |
Low-Pass Filter |
No |
|
Unique Features |
The Fluke 3000 FC Series Wireless Multimeter with the Fluke Connect™ app has all the essentials for convenient test and measurement troubleshooting:
AC and DC voltage measurements to 1000V.
AC and DC current with 0.01 mA resolution.
Continuity, resistance, diode test, capacitance and frequency measurements.
MIN/MAX recording.
CAT III 1000 V, Cat IV 600 V; IP54.
The Fluke a3001 FC Wireless iFlex® AC Current Module allows you to measure around awkward sized conductors or get into tight spaces and makes wire access easier:
Measure up to 2500 A AC.
Use as a standalone meter or as part of the system.
Logging function for recording and saving up to 65,000 readings.
Inrush function. |
Warranty |
1 YEARS |
Power Supply Voltage |
Battery Powered |
Safety Approval |
CSA, UL |
Data Logging |
Yes |
Interfaces I/O |
Bluetooth, Other, Wifi |
IP Rating |
IP54 |
Product Weight |
3.44 LBS |
Battery Type |
Other |
Product Height |
10 IN |
Product Length |
10.5 IN |
Product Width |
4.25 IN |
Shipping Weight |
3.78 LBS |
Country of Origin |
United States |
UPC |
095969724234 |
|