|
|
Additel ADT875-1210-AR-110V
Catalog #: ADT875-1210-AR-110V
Thermocouple Calibration Furnace
- Overview
- Specifications
- Summary
Product Highlights
- Temperature control from 100°C to 1210°C
- Two models to choose from: Reference (ADT878) and Standard (ADT875)
- Display Accuracy of ±1.5°C (ADT878)
- 4 on-board measurement channels (PC option)
- Process calibrator option provides a multi-channel readout for TCs, switches and transmitters, including task documentation and HART communication
- Portable, rugged and quick to temperature
- Self-calibration feature (PC option)
- Multi-zone temperature control
- Internal and external sensor control (PC option)
- Metallic interchangeable inserts
- Wi-Fi and Bluetooth capable
- Color touch screen display
- ISO 17025-accredited calibration w/data included
- Patent pending technology
Each Unit Includes
- Calibration Furnace and selected Insert & insulator
- Power adapter
- USB Cable
- Insert removal tool
- Accredited Calibration Certification
- Manual
|
875-1210 |
878-1210 |
Temperature Range |
100°C to 1210°C |
±1.2°C @ 100°C |
±1.0°C @ 100°C |
±1.2°C @ 300°C |
±1.0°C @ 300°C |
±1.2°C @ 600°C |
±1.0°C @ 600°C |
±1.6°C @ 900°C |
±1.2°C @ 900°C |
±2.0°C @ 1210°C |
±1.5°C @ 1210°C |
Stability |
±0.1°C |
Axial Uniformity (20mm zone) |
±0.6°C @ 100°C |
±0.4°C @ 100°C |
±0.9°C @ 300°C |
±0.6°C @ 300°C |
±1.2°C @ 600°C |
±0.8°C @ 600°C |
±1.5°C @ 900°C |
±1°C @ 900°C |
±1.5°C @ 1210°C |
±1°C @ 1210°C |
Radial Uniformity |
±0.2°C @ 100°C |
±0.2°C @ 100°C |
±0.3°C @ 300°C |
±0.3°C @ 300°C |
±0.4°C @ 600°C |
±0.4°C @ 600°C |
±0.8°C @ 900°C |
±0.6C @ 900°C |
±1°C @ 1210°C |
±0.8°C @ 1210°C |
Loading Effect |
±0.5°C |
Environmental Conditions |
8°C to 38°C guaranteed accuracy |
|
0°C to 50°C, 0% to 90% RH non-condensing, |
|
3000 M altitude for normal operation |
Storage Conditions |
-20°C to 60°C |
Immersion Depth |
XR style inserts = 138 mm (5.43") |
XS style inserts = 116 mm (4.57") |
(see insert ordering info for more details) |
|
Insert Size - OD |
24.8 mm (0.98 inches) |
Heating Time |
50 min: 23°C to 1210°C |
Cooling Time |
50 mins:1210°C to 300°C 50 mins: 300°C to 50°C |
55 mins:1210°C to 300°C 55 mins: 300°C to 50°C |
Typical Time to Stability |
15 min |
Resolution |
0.01°C |
Units |
°C, °F, and K |
Display |
6.5 in (165 mm) color touch screen |
Size (H x W x D) |
345 x 170 x 330 mm (13.6 x 6.7 x 13.0 in) |
Weight |
10.6 kg (23.4 lbs) |
Power Requirements |
115 V AC (±10%) or 230 V AC (±10%), switchable, 45-65 Hz, 580 W |
Mechanical Testing |
Vibration: 2 g (10-500 Hz) , 30 min for 2 sides |
Impact: 4 g three times |
Drop test: 500 mm (19.6 in) |
Communication |
USB A, USB B, RJ45, WiFi, Bluetooth |
Localization |
English, Chinese, Japanese, Russian, German |
Warranty |
1 year |
Input Specifications (Process Calibrator [PC] Option) |
|
875-1210 |
878-1210 |
TC Measurement Channels |
Patented TC terminals: |
|
Accepting S, R, K, B, N, E, J, T, L, and U |
TC Measurement Accuracy Type K Ch. 1-4 (excluding sensor) |
±0.182°C @ 100°C |
±0.172°C @ 100°C |
|
±0.182°C @ 100°C |
±0.172°C @ 100°C |
|
±0.266°C @ 300°C |
±0.236°C @ 300°C |
|
±0.310°C @ 600°C |
±0.251°C @ 600°C |
|
±0.397°C @ 900°C |
±0.304°C @ 900°C |
|
±0.517°C @1210°C |
±0.382°C @ 1210°C |
TC Range |
€“75 mV to 75 mV (UUT Channels 1-4) |
|
€“18 mV to 18 mV (Reference Channel) |
TC Resolution |
0.0001 mV, Input Impedance < 10Ω |
TC Voltage Accuracy |
0.02% RD + 8μV (ch. 1-4) |
0.01% RD + 8μV (ch. 1-4) |
|
0.01% RD + 2μV (ref ch.) |
0.005% RD + 2μV (Ref ch.) |
Internal CJC Accuracy |
±0.35°C (ch. 1-4) |
±0.30°C (ch. 1-4) |
|
±0.25°C (ref ch.) |
±0.20°C (ref ch.) |
Current Range |
€“30 mA to 30 mA |
Current Accuracy |
±(0.02% of rdg+ 2μA) |
±(0.01% of rdg + 0.6mV) |
0.001 V; Input impedance: > 1M 1MΩ |
0.0001V |
0.0001V |
DC 24V Output |
24 V ± 10%, MAX 60 mA |
Hart Communication |
Optional (ADT875PC and ADT878PC Models) |
"Temperature Coefficient 0°C to 8°C and 38°C to 50°C |
TC Readouts: ±5 ppm FS/°C |
|
Current: ±5 ppm FS/°C |
Switch Test |
Mechanical or Electrical - Channels 1 and 2 only |
Documentation |
Up to 1,000 tasks which store up to 10 results each containing as found and as left data. Snap shot feature allows for screen captures. Records auto step and ramp functions |
TC Measurement Specifications |
TC Type |
TEMP (°C) |
Error (°C)[1] |
875 |
875 |
K (CH1-CH4) |
100 |
±0.182 |
±0.172 |
300 |
±0.266 |
±0.236 |
600 |
±0.310 |
±0.251 |
900 |
±0.397 |
±0.304 |
1210 |
±0.517 |
±0.382 |
N (CH1- CH4) |
100 |
±0.273 |
±0.264 |
300 |
±0.270 |
±0.243 |
600 |
±0.309 |
±0.256 |
900 |
±0.368 |
±0.285 |
1210 |
±0.455 |
±0.335 |
E (CH1- CH4) |
100 |
±0.136 |
±0.126 |
300 |
±0.153 |
±0.130 |
600 |
±0.210 |
±0.154 |
900 |
±0.291 |
±0.202 |
1000 |
±0.297 |
±0.196 |
L (CH1- CH4) |
100 |
±0.223 |
±0.214 |
300 |
±0.271 |
±0.241 |
600 |
±0.308 |
±0.251 |
900 |
±0.522 |
±0.448 |
U (CH1- CH4) |
100 |
±0.270 |
±0.261 |
300 |
±0.189 |
±0.164 |
600 |
±0.227 |
±0.176 |
J (CH1- CH4) |
100 |
±0.186 |
±0.177 |
300 |
±0.197 |
±0.168 |
600 |
±0.256 |
±0.200 |
900 |
±0.281 |
±0.197 |
1200 |
±0.414 |
±0.294 |
S (CH1-CH4) |
100 |
±1.102 |
±1.094 |
300 |
±0.924 |
±0.899 |
600 |
±0.888 |
±0.837 |
900 |
±0.868 |
±0.793 |
1210 |
±0.865 |
±0.765 |
R (CH1- CH4) |
100 |
±1.080 |
±1.072 |
300 |
±0.869 |
±0.844 |
600 |
±0.804 |
±0.755 |
900 |
±0.771 |
±0.699 |
1210 |
±0.766 |
±0.670 |
B (CH1- CH4) |
250 |
±3.182 |
±3.170 |
300 |
±2.645 |
±2.631 |
600 |
±1.409 |
±1.379 |
900 |
±1.049 |
±1.003 |
1210 |
±0.905 |
±0.839 |
T (CH1- CH4) |
100 |
±0.194 |
±0.185 |
300 |
±0.191 |
±0.166 |
400 |
±0.217 |
±0.183 |
S (EXT. REF) |
100 |
±0.277 |
±0.273 |
300 |
±0.242 |
±0.229 |
600 |
±0.249 |
±0.224 |
900 |
±0.258 |
±0.220 |
1210 |
±0.266 |
±0.216 |
R (EXT. REF) |
100 |
±0.271 |
±0.266 |
300 |
±0.228 |
±0.216 |
600 |
±0.227 |
±0.202 |
900 |
±0.230 |
±0.194 |
1210 |
±0.240 |
±0.192 |
HTS/Schedule B Number |
9025.80.10 |
|
|
|
|