Fluke Calibration 7109A-156
Catalog #: 4757560
Portable Calibration Bath, -25 C TO 140 C, 115 VAC
- Overview
- Specifications
- Summary
- Accessories
Product Highlights
- Portable Bath, -25 C to 140 C, 115 VAC
- Calibrate up to four tri-clamp sanitary sensors at the same time
- Wide temperature ranges cover most process applications:
- 6109A: 35 °C to 250 °C
- 7109A: –25 °C to 140 °C
- Excellent display accuracy of ± 0.1 °C provides 4:1 test uncertainty ratio (TUR) for critical applications
- Easy to transport up stairs and across catwalks
- Stainless steel casing withstands harsh sterilizing chemicals and is rust proof
- Easy to use and maintain
- Global Fluke support and service
Each Unit Includes
- 7109A-156
- USB cable
- Probe access cover
- Transport tank cover
- Printed safety information
- Product CD (includes Operators Manual, USB Driver, Software distribution license, Source code for public domain software)
- NVLAP accredited calibration certificate
General specifications |
Size |
Height: 382 mm (15 in) Width: 242 mm (9.5 in) Depth: 400 mm (15.7 in) |
Weight |
6109A: 16 kg (35 lb) 7109A: 20 kg (44 lb) |
Fluid volume |
2.5 liters (0.66 gallons) |
Fluid working area |
75 mm x 75 mm (3 in x 3 in) |
Maximum fluid depth |
154 mm (6.1 in) |
Factory calibration |
Traceable NVLAP-accredited calibration included |
Temperature Specifications |
|
6109A |
7109A |
Range (at 25 °C ambient temperature) |
35 °C to 250 °C |
-25 °C to 140 °C (–15 °C cover off) |
Accuracy (maximum permissible error) |
0.1 °C |
0.1 °C |
Display resolution |
0.01°, 0.001° |
0.01°, 0.001° |
Stability |
0.015 °C |
0.01 °C |
Typical uniformity |
0.03 °C up to 200 °C
0.04 °C above 200 °C |
0.02 °C |
Repeatability |
0.04 °C |
0.04 °C |
Typical heating time |
35 °C to 100 °C: 25 minutes
100 °C to 250 °C: 45 minutes |
-25 °C to 25 °C: 35 minutes
25 °C to 140 °C: 55 minutes |
Typical cooling time |
250 °C to 100 °C: 35 minutes
100 °C to 35 °C: 55 minutes |
140 °C to 25 °C: 45 minutes
25 °C to –25 °C: 75 minutes |
Typical settling time |
15 minutes |
10 minutes |
Process Input Module Specifications (-P Models) |
Function |
Range |
Accuracy (Maximum Permissible Error) |
Reference input resistance |
0Ω to 42Ω
42 Ω to 400 Ω |
0.0025 Ω
0.006 % |
Reference input temperature |
-25 °C to 250 °C |
0.007 % + 0.015 °C |
Resistance sensing current |
1 mA |
8% |
DUT 4-wire resistance |
0 Ω to 31 Ω
31 Ω to 400 Ω |
0.0025 Ω
0.008 % |
DUT 3-wire resistance accuracy |
0 Ω to 400 Ω |
0.12 Ω |
Thermocouple mV |
-10 mV to 100 mV |
0.025 % + 0.01 mV |
Reference junction temperature |
0 °C to 40 °C |
0.35 °C |
Thermocouple temperature |
-25 °C to 250 °C |
J: 0.44 °C K: 0.49 °C T: 0.53 °C
E: 0.44 °C N: 0.57 °C M: 0.48 °C
L: 0.42 °C U: 0.48 °C |
0 °C to 250 °C |
R: 1.92 °C S: 1.88 °C C: 0.84 °C
D: 1.12 °C G/W: 3.34 °C |
mA range |
0 mA to 22 mA |
0.02 % + 0.002 mA |
Loop power |
24 V dc |
± 6 V |
ECCN Number |
EAR99 |
Country of Origin |
United States |
|
|
|
|