AEMC MX57EX TRMS
Catalog #: 2130.66
Digital Multimeter MX 57EX TRMS (IECEx LCIE 07.0010 X, LCIE 02 ATEX 6005 X, Intrinsically safe, TRMS, 50,000-count, 0.025% Accuracy)
- Overview
- Specifications
- Summary
- Current Clamps/Probes
- Accessories
Discontinued
Product Highlights
- LCI 07: 0010X
- Safety rating: IEC 61010-1: 2001
- Agency Approval: IECEx LC107.0010 X, LCIE 02 ATEX 6005 X, Ex II 2 G D or Ex I M2, Ex ib I, Ex ib IIC T5 or T4 or T3 ,Ex ibD 21 T
- Logic signal measurement and ADP input
- Min/Max/Avg functions
- Bargraph with zoom (x5) and center zero
- Rugged design — IP67 rating
- Protection by 500mA intrinsic safety fuse for the current range
- Includes test leads
Applications:
- Oil refineries
- Mining
- Pharmaceutical plants
Each Unit Includes
- Test Leads
- 9V Battery (alkaline)
AC Current |
500μA |
Resolution |
10nA |
Bandwidth |
DC to 5kHz |
Accuracy |
±0.75% of Reading ± 30cts |
Protection |
600Vrms |
5mA |
Resolution |
100nA |
Bandwidth |
DC to 5kHz |
Accuracy |
±0.6% of Reading ± 30cts |
Protection |
600Vrms |
50mA |
Resolution |
1μA |
Bandwidth |
DC to 5kHz |
Accuracy |
±0.6% of Reading ± 30cts |
Protection |
600Vrms |
500mA |
Resolution |
10μA |
Bandwidth |
DC to 5kHz |
Accuracy |
±0.7% of Reading ± 30cts |
Protection |
600Vrms |
AC Voltage |
500mV |
Resolution |
10μV |
Bandwidth |
40Hz to 1kHz |
Accuracy |
±0.3% of Reading ± 30cts |
Input Impedance |
11MΩ |
Protection |
1100Vpk |
5V |
Resolution |
100μV |
Bandwidth |
1kHz to 4kHz |
Accuracy |
±1% of Reading ± 30cts |
Input Impedance |
11MΩ |
Protection |
1100Vpk |
50V |
Resolution |
1mV |
Bandwidth |
4kHz to 10kHz |
Accuracy |
±2% of Reading ± 30cts |
Input Impedance |
10MΩ |
Protection |
1100Vpk |
500V* |
Resolution |
10mV |
Bandwidth |
10kHz to 30kHz |
Accuracy |
±2% of Reading ± 30cts |
Input Impedance |
10MΩ |
Protection |
1100Vpk |
1000V* |
Resolution |
100mV |
Bandwidth |
30kHz to 50kHz |
Accuracy |
±3% of Reading ± 30cts |
Input Impedance |
10MΩ |
Protection |
1100Vpk |
DC Current |
500μA |
Resolution |
10nA |
Accuracy |
±0.2% of Reading ± 5cts |
Protection |
600Vrms |
5mA |
Resolution |
100nA |
Accuracy |
±0.2% of Reading ± 2cts |
Protection |
600Vrms |
50mA |
Resolution |
1μA |
Accuracy |
±0.05% of Reading ± 2cts |
Protection |
600Vrms |
500mA |
Resolution |
10μA |
Accuracy |
±0.2% of Reading ± 2cts |
Protection |
600Vrms |
DC Voltage |
500mV |
Resolution |
10μV |
Accuracy |
±0.025% of Reading ± 2cts |
Input Impedance |
11MΩ |
Protection |
1100Vpk |
5V |
Resolution |
100μV |
Accuracy |
±0.025% of Reading ± 2cts |
Input Impedance |
11MΩ |
Protection |
1100Vpk |
50V |
Resolution |
1mV |
Accuracy |
±0.025% of Reading ± 2cts |
Input Impedance |
10MΩ |
Protection |
1100Vpk |
500V* |
Resolution |
10mV |
Accuracy |
±0.025% of Reading ± 2cts |
Input Impedance |
10MΩ |
Protection |
1100Vpk |
1000V* |
Resolution |
100mV |
Accuracy |
±0.2% of Reading ± 2cts |
Input Impedance |
10MΩ |
Protection |
1100Vpk |
Resistance |
500Ω |
Resolution |
10mΩ |
Accuracy |
±0.07% of Reading ± 5cts |
Max Open-Circuit Voltage |
7V |
Protection |
600V |
5kΩ |
Resolution |
100mΩ |
Accuracy |
±0.07% of Reading ± 2cts |
Max Open-Circuit Voltage |
7V |
Protection |
600V |
50kΩ |
Resolution |
1Ω |
Accuracy |
±0.07% of Reading ± 2cts |
Max Open-Circuit Voltage |
7V |
Protection |
600V |
500kΩ |
Resolution |
10Ω |
Accuracy |
±0.07% of Reading ± 2cts |
Max Open-Circuit Voltage |
7V |
Protection |
600V |
5MΩ |
Resolution |
100Ω |
Accuracy |
±0.3% of Reading ± 2cts |
Max Open-Circuit Voltage |
7V |
Protection |
600V |
50MΩ |
Resolution |
1kΩ |
Accuracy |
±1% of Reading ± 2cts |
Max Open-Circuit Voltage |
7V |
Protection |
600V |
|
Continuity |
10 to 20O |
Response Time |
1ms |
Diode Test |
0 to 2V |
Test Current |
1mA ± 20% |
Capacitance |
50nF to 50mF |
Accuracy |
1% of Reading ± 2cts |
Frequency |
0.62Hz to 500kHz |
Accuracy |
0.03% of Reading ± 2cts |
Temperature Range
(User selectable in °F or °C) |
-328° to 1472°F (-200° to 800°C) |
Temperature Sensor |
PT100/PT1000 |
Digital Display |
50,000-count |
Analog Bargraph |
34-segment |
Safety Rating |
NF EN 50014, EN 50014: 1992; NF EN 50020, EN 50020: 1994 |
Agency Approval |
Agency Approval: IECEx LC107.0010 X, LCIE 02 ATEX 6005 X, ExII 2 G D or ExI M2, Ex ib I, Ex ib IIC T5 or T4 or T3, Ex ibD 21 T |
Multimeters Template |
Type |
Digital |
Style |
Hand-Held |
Measures AC V |
Yes |
Measures DC V |
Yes |
Measures AC A |
Yes |
Measures DC A |
Yes |
True RMS |
Yes |
Maximum Voltage AC |
1000 V |
Minimum Voltage AC |
10 uV (0.01 mV) |
Maximum Voltage DC |
1000 V |
Minimum Voltage DC |
10 mV |
Maximum Current AC |
500 mA (0.5 A) |
Minimum Current AC |
10 nA (0.00001 mA) |
Maximum Current DC |
500 mA (0.5 A) |
Minimum Current DC |
10 mA |
Basic Accuracy V DC |
±0.025%of reading ± 2cts |
# Digit Display |
4 |
Backlight |
Yes |
Manual/Autoranging |
Both |
Dual Display |
No |
Analog Bargraph on Display |
Yes |
Decibels |
No |
Resistance |
Yes |
Frequency |
Yes |
Capacitance |
Yes |
Temperature |
Yes |
Diode Testing |
Yes |
CAT Rating |
III |
Intrinsically Safe |
Yes |
Kelvin 4-Wire Measurement Capability |
No |
Wireless Display |
No |
Data Hold |
Yes |
Low-Pass Filter |
Yes |
Data Logging |
No |
Warranty |
1 YEARS |
Calibration Included |
Factory Calibration |
IP Rating |
IP64 |
Product Weight |
400 GRAMS (0.882 LBS) |
Product Height |
40 MM (1.57 IN) |
Product Length |
189 MM (7.44 IN) |
Product Width |
82 MM (3.23 IN) |
ECCN Number |
EAR99 |
Country of Origin |
France |
|
Current Clamps/Probes
Product |
Price |
|
AEMC MN213
Catalog #: 2115.75
AC Current Probe Model MN213 (Lead - 1mA/A - 240A max)
|
$157.21
|
|
AEMC MN373
Catalog #: 2116.28
AC Current Probe Model MN373 (Lead - 1000mV/A - 2.4A max & 10mV/A 150A max)
|
$230.21
|
|
AEMC SL261
Catalog #: 1201.51
AC/DC Current Probe Model SL261 (BNC - 100mV/A - 10A peak & 10mV/A - 100A peak)
|
$459.21
|
|
AEMC SR604
Catalog #: 2113.44
AC Current Probe Model SR604 (1000A, 1mA/A, Lead) L**
|
$214.21
|
Accessories
Product |
Price |
|
AEMC 2970.92
Catalog #: 2970.92
Fuse Set of 5, 1100V, 11A, 10x38mm, 20kA for Models MTX 3281B-3283B-BT
|