Fluke Calibration 6532-140M
Catalog #: 4183692
Electronic Deadweight Tester Kits, 140 MPA (20K PSI), EXTENDED RANGE
- Overview
- Specifications
- Summary
Product Highlights
- E-DWT-H A140M/A14M Configuration
- Pressure range 1.4 to 140 MPa, 200 to 20,000 psi
- ± 0.02 % of reading total one-year accuracy, 0.025% for two years
- Low torque variable volume allows for pressure generation and control up to 200 MPa (30,000 psi) with minimal physical effort
- Built-in priming system to fill system with test fluid and purge unwanted air to assure smooth operation
- Unique test port design eliminates the need for PTFE tape or wrenches. Adapters included for 1/8 in, 1/4 in, 3/8 in and 1/2 in NPT, 1/8 in, 1/4 in, 3/8 in and 1/2 in BSP, M20 x 1.5 and M14 x 1.5
- Compatible with a wide range of liquids, including water. Vacuum fill kit included to fill the E-DWT with any compatible test medium
- AutoTest, AutoRange, ready/not ready indicator, onboard data capture and storage, overpressure alarm, RS-232 remote interface and other advanced features to simplify testing, improve safety and prevent calibration errors and damage to equipment
- Optional foot switch accessory allows hands-free data collection when running AutoTests
- Rechargeable battery pack and transport case options for field operation
Each Unit Includes
- E-DWT-H Electronic Deadweight Tester with power supply and operating manual
- Vacuum liquid fill kit
- Test station quick-mount fitting
- NPT adapters: 1/8 in, 1/4 in, 3/8 in and 1/2 in
- BSP adapters: 1/8 in, 1/4 in, 3/8 in and 1/2 in
- Metric adapters: M20 x 1.5 and M14 x 1.5
General |
Power requirements |
To RPM4-E-DWT: 12 V dc 1.2 A
To ac to dc power supply: 100 V ac to 240 V ac, 50/60 Hz |
Operating temperature range |
10 °C to 40 °C (50 °F to 104 °F) |
Storage temperature range |
-20 °C to 70 °C (-4 °F to 158 °F) (must be above freezing point and below boiling point of test medium) |
Relative operating humidity |
0 % to 70 % |
Relative storage humidity |
0 % to 100 % |
Weight |
E-DWT with 1 Q-RPT: 12 kg (26 lb) approximate
E-DWT with 2 Q-RPTs: 14 kg (30 lb) approximate |
Dimensions |
E-DWT footprint: 41.4 cm W x 37.1 cm D (16.3 in x 14.6 in)
E-DWT height: 26.9 cm (10.6 in), 33.6 cm (13.2 in) to max variable volume handle height |
Pressure range |
Up to 200 MPa (30k psi), depending on configuration |
Operating medium |
All 6531 and 6532 E-DWT Kits are shipped dry, standard preparation. Can be filled with di-ethyl-hexyl sebacate, silicon oils, propylene glycol, fully fluorinated liquids, partially fluorinated liquids, isopropyl alcohol and distilled water or mineral oil. |
Reservoir capacity |
300 cc (18 in 3 ) |
Variable volume displacement |
3 cc (0.18 in 3 ), 200 MPa (30 000 psi) maximum |
Filling and priming pump displacement |
3.7 cc (0.23 in 3 ) |
TEST pressure connection |
DH500 female, 200 MPa (30k psi) maximum working pressure. Adapters included for 1/8 in, 1/4 in, 3/8 in and 1/2 in NPT, 1/8 in, 1/4 in, 3/8 in and 1/2 in BSP, M20 x 1.5 and M14 x 1.5. Adapter maximum working pressure is 140 MPa (20k psi). Note: DH500 is a gland and collar type fitting for 6 mm (1/4 in) coned and left hand threaded tubes equivalent to AE F250C, HIP HF4, 9/16-18 UNF, etc. |
Pressure limits |
Maximum working pressure: Range of RPM4-E-DWT monitor’s Hi Q-RPT
Maximum priming pump pressure: 700 kPa (100 psi)
Maximum working pressure with Lo Q-RPT selected: Range of RPM4-E-DWT monitor’s Lo Q-RPT |
Communication ports |
RS-232 (COM1, COM2) |
Pressure measurement |
Warm up time |
15 minute temperature stabilization recommended from cold power up |
Normal operating temperature range |
10 °C to 40 °C (50 °F to 104 °F) |
Resolution |
Default: 0.01 % of active range
User adjustable to 1 ppm of Q-RPT maximum or 10 ppm of active AutoRange, whichever is larger |
Precision 1 |
± 0.018 % of reading or 0.0018 % of Q-RPT span, whichever is greater |
Predicted Stability 2 |
One year: ± 0.0075 % of reading
Two year: ± 0.015 % of reading |
Measurement uncertainty 3 |
One year: ± 0.02 % of reading or 0.002 % of Q-RPT span, whichever is greater
Two year: ± 0.025 % of reading or 0.0025 % of Q-RPT span, whichever is greater |
1. Combined linearity, hysteresis, and repeatability. Precision does not include stability or calibration reference uncertainty.
2. Predicted Q-RPT measurement stability limit (k=2) assuming regular use of AutoZero function and short term stability between rezeroing.
3. Maximum deviation of the Q-RPT indication from the true value of applied pressure including precision, predicted stability with rezeroing, temperature effect from 10 ºC to 40 ºC and calibration uncertainty (assumes calibration reference uncertainty of ± 0.005 % of reading, k=2), combined and expanded (k=2) following the ISO “Guide to the Expression of Uncertainty in Measurement”. |
ECCN Number |
EAR99 |
Country of Origin |
United States |
|
|
|
|