Fluke Calibration P3031-KGCM2
Catalog #: 4184783
Gas Deadweight Tester Liquid Lubricated, 70 KGF/CM2
- Overview
- Specifications
- Summary
- software
- Accessories
Product Highlights
- Liquid Lubricated - Gas Operated with optional hand pumps
- Pressure Range 1 to 70 kgf/cm
- Accuracy better than 0.015 % of reading. (Increased accuracy option of 0.008 %)
- Dual vacuum/pressure models available
- Psi, bar, kgf/cm 2 , kPa and MPa ranges available
- Piston/cylinder design provides stability and repeatability
- Built-in pneumatic hand pumps for pressure and vacuum
- High quality needle valves provide optimum control
- Mounted spirit level and adjustable feet
- Test station design with O-rings eliminates the need for PTFE tape or wrenches
- Improved case with spring-loaded latches for hood
- Sturdy weight box with hinged lid and side handles for easy transport
Specifications |
Accuracy |
±0.015% of reading (±0.008% optional) |
*Accuracy based on % of reading from 10% to 100% of the piston range when used in accordance with the corrections found on the calibration certificate. Below 10% ± (accuracy class) x 10% of the piston range. |
Mass and PCU Materials of Construction |
Standard weight material |
Series 3 non-magnetic austenitic stainless steel |
Weight density |
2.7 g/cm³ |
Optional fractional weights |
Solution heat treated aluminum |
Models P3011, P3012, P3013, P3022, P3023, P3025 (V) |
Piston Material:
Cylinder Material:
Coefficient of Expansion: |
Ceramic
Martensitic steel
11 ppm/°C |
Models P3014, P3015, P3025 (P) P3031, P3032 (P & V) |
Piston Material:
Cylinder Material:
Coefficient of Expansion: |
Tungsten carbide
Martensitic steel
16.5 ppm/°C |
Model P3016 |
Piston Material:
Cylinder Material:
Coefficient of Expansion: |
Tungsten carbide
Tungsten carbide
11 ppm/°C |
General |
Test port adaptors |
1/8, 1/4, 3/8 and ½ NPT or BSP |
Instrument weight |
24 lbs |
Instrument size |
17.5 W x 12 D x 8.5 H |
Mass set weight (typical) |
65 lbs |
Minimum Standard Weight Increments |
P3012, P3022 Pressure |
1 inH2O, (5 mbar) |
P3013, P3023 Pressure |
2 inH2O, (10 mbar) |
P3014, P3015, P3025 Pressure |
1 psi, (0.1bar) |
P3016, P3031, P3032 Pressure |
1 psi, (0.1bar) |
P3011, P3022, P3023, P3025 Vacuum |
0.2 inHg (10 mbar) |
Optional Fractional Weights |
P3014, P3015, P3025 Pressure |
0.1 psi (0.01 bar) |
Built-in Hand Pumps |
Pressure mode |
Max pressure 300 psi (20 bar) |
Vacuum mode |
To 90% vacuum |
Operating Fluid for Liquid Lubricated Models |
55-655 |
mineral oil (standard) |
Krytox |
for oxygen-safe applications (optional) |
Options |
PressCal Software |
Windows-based software program that allows users to easily apply all necessary corrections to enhance the deadweight tester performance. Calibration details are then stored and/or used to automatically create a calibration certificate. PressCal is provided as standard with all 0.008% instruments. |
Krytox |
As operating fluid for liquid lubricated gas operated models. |
ECCN Number |
EAR99 |
Country of Origin |
United States |
|
|
|
|