Fluke 750R31
Catalog #: 4352691
Pressure Modules, 0 TO 10000 PSI (0 TO 700 BAR), (0 TO 700 MPA)
- Overview
- Specifications
- Summary
- Calibration Pumps
- Accessories
Product Highlights
- 750R31 Pressure Module
- 0 psi to 10000 psi (0 to 700 bar)
- (0 to 70 MPa)
- 0.025 % reference uncertainty
- 6-month and 1-year specifications
- Temperature compensated 0 °C to 50 °C
- Digital communication to calibrators, no analog losses or errors
- Broad selection of ranges
- Gage, differential, dual range, absolute and vacuum measurement models
Each Unit Includes
- 1/8 in. NPT Male-to-1/4 in. NPT Male Adapter
- 1/8 in. NPT Male-to-1/4 BSP Male Adapter
- 1/8 in. NPT Male-to-M20 Male Adapter
- Certificate of Calibration
- O-RING (-111) Cast Urethane
- User Manual
Reference Module Specifications |
Model |
Parameter / Range |
Burst Rating6 |
Hi Side Media2 |
Lo Side Media2 |
750RD6 |
-12 to 100 psi (-1 to 7 bar) |
4X |
316 SS |
— |
750R27 |
0 to 300 psi (0 to 20 bar) |
4X |
316 SS |
— |
750R29 |
0 to 3000 psi (0 to 200 bar) |
3X |
316 SS |
— |
750R31 |
0 to 10000 psi (0 to 700 bar) |
2X |
316 SS |
— |
750RD27 |
-12 to 300 psi (-0.8 to 20 bar) |
4X |
316 SS |
— |
Model |
Reference
Uncertainty4 |
Total Uncertainty
1-year
(15-35°C) |
Total Uncertainty
1-year1 |
Total Uncertainty
6-month
(15-35°C) |
Total Uncertainty
6-month1 |
750RD6 |
±0.01% of FS |
±0.02% of FS |
±0.04% of FS |
±0.015% of FS |
±0.035% of FS |
750R27 |
±0.01% of FS |
±0.02% of FS |
±0.04% of FS |
±0.015% of FS |
±0.035% of FS |
750R29 |
±0.01% of FS |
±0.02% of FS |
±0.04% of FS |
±0.015% of FS |
±0.035% of FS |
750R31 |
±0.01% of FS |
±0.02% of FS |
±0.04% of FS |
±0.015% of FS |
±0.035% of FS |
750RD27 |
±0.01% of FS |
±0.02% of FS |
±0.04% of FS |
±0.015% of FS |
±0.035% of FS |
1. Total uncertainty, % of full span for temperature range 0°C to +50°C, one year interval. Total uncertainty,
1.0% of full span for temperature range -10°C to 0°C, one year interval. No 6 month specification available for range -10°C to 0°C.
2. "Non-corrosive gasses" indicates dry air or non-corrosive gas as compatible media. "Stainless Steel 316-SS"
indicates media compatible with Type 316 Stainless Steel.
3. Specifications % of Full Span unless otherwise noted.
4. Reference Uncertainty is the specification for as left data for 24 hours.
5. When reference class modules are used with fixed resolution products (717, 718, 719 series, 725 and 726)
calibrators add ±1 count to the overall accuracy specification.
6. Burst rating specification refers to the multiplier times full scale of the module for the rated burst
pressure. |
General Specifications |
Mechanical Specification |
Size (H X W X L) |
45 mm X 94 mm X 110 mm
(1.77 in X 3.70 in X 4.33 in) |
Weight |
.292 g (10.3 oz) |
Physical Interface |
Serial Connector, Pressure Port |
Environmental Specification |
OperatingTemperature |
-10 °C to +50 °C (14 °F to 122 °F) |
Storage Temperature |
-20 °C to +60 °C (-4 °F to 140 °C) |
Operating Humidity |
Non condensing (<10 °C) (<50 °F)
90 % RH (10 °C to 30 °C) (50 °F to
86 °F)
75 % RH (30 °C to 40 °C) (86 °F to
104 °F)
45 % RH (40 °C to 50 °C) (104 °F
to 122 °C) |
Operating Altitude |
2000 m (6,561 ft) |
Storage Altitude |
12000 m (45,700 ft) |
IP Rating |
IP 52 |
Compliance |
CE |
IEC 61010-1
Protection Class II
Pollution Degree 2 |
EMI, RFI, EMC |
.EN61326-1
Electromagnetic Compatibility
Applies to use in Korea only.
Class A equipment (Industrial
Broadcasting & Communication
Equipment) [1]
[1] This Product meets
requirements for industrial (Class A)
electromagnetic wave equipment
and the seller or user should
take notice of it. This equipment
is intended for use in business
environments and is not to be used
in homes. |
Unique Features |
0 psi to 10000 psi (0 to 700 bar), (0 to 70 MPa) |
Product Weight |
10.3 OUNCES (0.644 LBS) |
Product Height |
1.77 IN |
Product Length |
4.33 IN |
Product Width |
3.70 IN |
Country of Origin |
China |
|
Calibration Pumps