Fluke Calibration 1504-156
Catalog #: 1648966
Tweener Thermistor Readout, 110V 50/60HZ
- Overview
- Specifications
- Summary
- Software
- Accessories
- Cases
- Temp. Probes
Product Highlights
- Two Tweeners to choose from—reading PRTs or thermistors
- Battery packs available
- Best price/performance package
Each Unit Includes
- 1504 Thermometer
- Extra Probe Connector
- Power Cord
- Manual
- Probe (optional—must be purchased separately)
- Battery Pack (optional—must be purchased separately)
Specifications |
|
1502A |
1504 |
Temperature Range† |
-200 °C to 962 °C
(-328 °F to 1764 °F) |
Any thermistor range |
Resistance Range |
0 Ω to 400 Ω, auto-ranging |
0 Ω to 1 MΩ, auto-ranging |
Probe |
Nominal RTPW: 10 Ω to 100 Ω RTD, PRT, or SPRT |
Thermistors |
Characterizations |
Complies with European Union Directive 89/336/EEC, EN61326-1. |
SAFETY: |
ITS-90 subranges 4, 6, 7, 8, 9, 10, and 11
IPTS-68: R0, a, d, a4, and c4
Callendar-Van Dusen: R0, a, d, and b |
Steinhart-Hart thermistor polynomial
Callendar-Van Dusen: R0, a, d, and b |
Resistance Accuracy (ppm of reading) |
0 Ω to 20 Ω: 0.0005 Ω
20 Ω to 400 Ω: 25 ppm |
0 Ω to 5 KΩ: 0.5 Ω
5 KΩ to 200 KΩ: 100 ppm
200 KΩ to 1 MΩ: 300 ppm |
Temperature Accuracy† |
±0.004 °C at -100 °C
±0.006 °C at 0 °C
±0.009 °C at 100 °C
±0.012 °C at 200 °C
±0.018 °C at 400 °C
±0.024 °C at 600 °C |
±0.002 °C at 0 °C
±0.002 °C at 25 °C
±0.004 °C at 50 °C
±0.010 °C at 75 °C
±0.020 °C at 100 °C
(Using 10 KW thermistor sensor, a=0.04. Does not include probe uncertainty or characterization errors.) |
Operating Temperature Range |
16 °C to 30 °C |
13 °C to 33 °C |
Resistance Resolution |
0 Ω to 20 Ω: 0.0001 Ω
20 Ω to 400 Ω: 0.001 Ω |
0 Ω to 10 KΩ: 0.01 Ω
10 KΩ to 100 KΩ: 0.1 Ω
100 KΩ to 1 MΩ: 1 Ω |
Temperature Resolution |
0.001 °C |
0.0001 °C |
Excitation Current |
0.5 and 1 mA, user selectable, 2 Hz |
2 and 10 µA, automatically selected |
Measurement Period |
1 second |
Digital Filter |
Exponential, 0 to 60 seconds time constant (user selectable) |
Probe Connection |
4-wire with shield, 5-pin DIN connector |
Communications |
RS-232 serial standard
IEEE-488 (GPIB) optional |
Display |
8-digit, 7-segment, yellow-green LED; 0.5-inch-high characters |
Power |
115 VAC (±10 %), 50/60 Hz, 1 A, nominal
230 VAC (±10 %), 50/60 Hz, 1 A, nominal, specify |
Size (HxWxD) |
61 x 143 x 181mm (2.4 x 5.6 x 7.1 in) |
Weight |
1.0 kg (2.2 lb.) |
Probes from Fluke Calibration |
5615, 5627, 5626, 5628, 5622 |
5640-44, 5610-65 |
Calibration |
Accredited NIST-traceable calibration provided |
†Temperature ranges and accuracy may be limited by the sensor you use.
|
Type (Contact Thermometers) |
RTD |
Calibration Included |
NIST Calibration |
Power Supply Voltage |
120V 50/60 Hz |
Safety Approval |
CSA |
Interfaces I/O |
GPIB, Other, RS-232 |
Product Weight |
2.2 LBS |
Product Height |
181 MM (7.13 IN) |
Product Length |
61 MM (2.40 IN) |
Product Width |
143 MM (5.63 IN) |
Country of Origin |
United States |
|
Software