Fluke Calibration 5606-50-A
Catalog #: 3622827
Full Immersion PRT Probe, 50MM, -200 TO 160C, W/O CAL, A = INFO-CON for 914X
- Overview
- Specifications
- Summary
- Cases
Product Highlights
- 5606-50-A = “A” (INFO-CON for 914X)
- Temperature range from -200 °C to 160 °C
- Calibration accuracy of ± 0.05 °C
- Transition junction and lead wires can withstand full PRT temperature range
- Unique probe seal of 5606 prevents ingress of moisture
- Calibration not included; NVLAP-accredited calibration optional
Each Unit Includes
- Coefficient Sheet
- User Manual CD
- INFO-CON Addendum
|
|
Temperature range |
-200 °C to 160 °C 0 °C to 450 °C |
Nominal resistance at 0.01 °C |
100 O ± 0.1 O |
Temperature coefficient |
0.00385 O/O/°C |
Accuracy |
± 0.04 °C at 0.01 °C
± 0.06 °C at 160 °C |
± 0.04 °C at 0.01 °C
± 0.07 °C at 450 °C |
Short-term repeatability |
± 0.03 °C at 0.01 °C
± 0.04 °C at 160 °C |
± 0.020 °C at 0.01 °C
± 0.035 °C at 450 °C |
Drift |
± 0.03 °C at 0.01 °C
± 0.04 °C at 160 °C |
± 0.02 °C at 0.01 °C
± 0.04 °C at 450 °C |
Hysteresis |
± 0.015 °C |
Sheath length |
50 mm ± 5 mm
(2 in ± 0.2 in) |
40 mm ± 5 mm
(1.6 in ± 0.2 in) |
Sheath diameter |
3.1 mm ± 0.1 mm (1/8 in ± 0.004 in) |
Sheath material |
316 SST |
Inconel ± 600 |
Transition junction temperature range |
-200 °C to 160 °C |
0 °C to 450 °C |
Transition junction dimensions |
No transition |
35 mm x 7 mm (1.4 in x 0.3 in) |
Sensor length |
30 mm ± 3mm (1.2 in ±0.1 in) |
Sensor location |
3 mm ± 1 mm from tip (1.2 in) |
Minimum insulation resistance |
20 MO at 23 °C |
100 MO at 23 °C |
Minimum immersion length |
Full immersion |
Full immersion |
Maximum immersion depth in liquid medium |
Full immersion |
Use SST or glass tube to protect transition junction |
Maximum immersion depth in dry medium |
Full immersion |
Full immersion |
Response time |
12 seconds typical |
9 seconds typical |
Self heating (in 0 °C bath) |
± 0.003 °C |
Lead-wire cable type |
Enameled copper wire |
SST over-braid |
Lead-wire length |
2.4 meters (8 ft) |
Lead-wire temperature range |
160 °C |
450 °C |
Calibration |
Calibration not included. NVLAP accredited calibration available. See ordering information. |
[1] Includes calibration and 100 hr drift (k = 2).
[2] Three thermal cycles from min to max temp, includes hysteresis, 95 % confidence (k = 2).
[3] After 100 hours at max temp, 95 % confidence (k = 2).
[4] Temperatures outside this range will cause irreparable damage.
[5] Per ASTM E 644. |
Fluke Calibration Platinum Resistance Thermometers (PRTs) |
Model |
Range |
Size |
Basic Accuracy† |
Platinum Resistance Thermometer: Secondary Standard PRT |
5608-9-X |
–200 to 500 °C |
9 in x 1/8 in |
Select from available calibration options |
5608-12-X |
–200 to 500 °C |
12 in x 1/8 in |
5609-12-X |
–200 to 670 °C |
12 in x 1/4 in |
5609-15-X |
–200 to 670 °C |
15 in x 1/4 in |
5609-20-X |
–200 to 670 °C |
20 in x 1/4 in |
5609-300-X |
–200 to 670 °C |
300 mm x 6 mm |
5609-400-X |
–200 to 670 °C |
400 mm x 6 mm |
5609-500-X |
–200 to 670 °C |
500 mm x 6 mm |
5626 |
–200 °C to 661 °C |
305 or 381 x 6.35 mm (12 or 15 x 0.25 in |
±0.007 °C at 0 °C |
Platinum Resistance Thermometer: Secondary Reference PRT |
5616-12 |
–200 °C to 420 °C |
6.35 mm x 298 mm (0.250 x 11.75 in) |
± 0.010 °C at 0.010 °C |
5615-6 |
–200 to 300 °C |
4.76 mm x 152 mm (0.188 x 6.0 in) |
± 0.013 °C at 0.010 °C |
5615-9 |
–200 °C to 420 °C |
4.76 mm x 229 mm (0.188 x 9.0 in) |
± 0.013 °C at 0.010 °C |
>5615-12 |
–200 °C to 420 °C |
6.35 mm x 305 mm (0.250 x 12.0 in) |
± 0.013 °C at 0.010 °C |
Platinum Resistance Thermometer: Precision Industrial PRT |
5627A-6 |
–200 °C to 300 °C |
152 x 4.7 mm (6 x 0.187 in) |
±0.05 °C at 0 °C |
5627A-9 |
–200 °C to 300 °C |
229 x 4.7 mm (9 x 0.187 in) |
±0.05 °C at 0 °C |
5627A-12 |
–200 °C to 420 °C |
305 x 6.35 mm (12 x 0.25 in) |
±0.05 °C at 0 °C |
Platinum Resistance Thermometer: Fast Response PRT |
>5622-05 |
–200 °C to 350 °C |
100 x 0.5 mm |
±0.04 °C at 0 °C |
>5622-10 |
–200 °C to 350 °C |
100 x 1.0 mm |
±0.04 °C at 0 °C |
>5622-16 |
–200 °C to 350 °C |
200 x 1.6 mm |
±0.04 °C at 0 °C |
>5622-32 |
–200 °C to 350 °C |
200 x 3.2 mm |
±0.04 °C at 0 °C |
Platinum Resistance Thermometer: Small Diameter Industrial PRTs |
5618B-6 |
–200 °C to 300 °C |
152 x 3.2 mm (6 x 0.125 in) |
±0.05 °C |
5618B-9 |
–200 °C to 500 °C |
229 x 3.2 mm (9 x 0.125 in) |
±0.05 °C |
5618B-12 |
–200 °C to 500 °C |
305 x 3.2 mm (12 x 0.125 in) |
±0.05 °C |
Platinum Resistance Thermometer: Full Immersion PRTs |
5606 Immersion PRT |
–200 to 160 °C |
50 x 3.1 mm ( 2 x 1/8 in) |
±0.05 °C |
5607 Immersion PRT |
0 to 450 °C |
75 x 3.1 mm (3 x 1/8 in) |
±0.055 °C |
5623B Freezer Probe |
–100 °C to 156 °C |
152 x 6.35 mm (6 x 0.25 in) |
±0.05 °C |
Thermistors |
Thermistor: Standards |
5640 |
0 °C to 60 °C |
229 x 6.35 mm (9 x 0.25 in) |
±0.0015 °C |
5641 |
0 °C to 60 °C |
114 x 3.2 mm (4.5 x 0.125 in) |
±0.001 °C |
5642 |
0 °C to 60 °C |
229 x 3.2 mm (9 x 0.125 in) |
±0.001 °C |
5643 |
0 °C to 100 °C |
114 x 3.2 mm (4.5 x 0.125 in) |
±0.0025 °C |
5644 |
0 °C to 100 °C |
229 x 3.2 mm (9 x 0.125 in) |
±0.0025 °C |
Thermistor: Secondary Probes |
5610 |
0 °C to 100 °C |
152 or 229 x 3.2 mm (6 or 9 x 0.125 in) |
±0.01 °C |
5611A |
0 °C to 100 °C |
1.5 mm (0.06 in) tip dia. |
±0.01 °C |
5611T |
0 °C to 100 °C |
28 x 3 mm (1.1 x 0.12 in) |
±0.01 °C |
†“Basic Accuracy” includes calibration uncertainty and short-term repeatability. It does not include long-term drift.
X = termination. Specify A (INFO-CON for 914X), B (bare wire), D (5-pin DIN for Tweener Thermometers), G (gold pins), I (INFO-CON for 1521 or 1522), J (banana plugs), L (mini spade lugs), M (mini banana plugs), P (INFO-CON for 1523 or 1524), or S (spade lugs). |
Accessories/Probes/Temperature Probes Template |
Application/Properties |
Immersion |
Max Temperature |
160 C |
Min Temperature |
-200 F |
Wire Termination |
Other |
Insulation Material |
Stainless Steel |
Probe Length |
50 IN |
Lead Length |
96 IN (25.40 MM) |
Unique Features |
Temperature range –200 °C to 160 °C
Nominal resistance at 0.01 °C 100 Ω ± 0.1 Ω
Temperature coefficient 0.00385 Ω/Ω/°C
Accuracy ± 0.04 °C at 0.01 °C
± 0.06 °C at 160 °C
Short-term repeatability ± 0.03 °C at 0.01 °C
± 0.04 °C at 160 °C
Drift ± 0.03 °C at 0.01 °C
± 0.04 °C at 160 °C
Hysteresis ± 0.015 °C
Sheath length 50 mm ± 5 mm
(2 in ± 0.2 in)
Sheath diameter 3.1 mm ± 0.1 mm (1/8 in ± 0.004 in)
Sheath material 316 SST
Transition junction temperature range[4] –200 °C to 160 °C |
Warranty |
1 YEARS |
Additional Info/Comments |
Sensor length 30 mm ± 3mm (1.2 in ±0.1 in)
Sensor location 3 mm ± 1 mm from tip (1.2 in ± 0.04 in)
Minimum insulation resistance 20 MΩ at 23 °C
Minimum immersion length Full immersion
Maximum immersion depth in liquid medium Full immersion
Response time 12 seconds typical
Self heating (in 0 °C bath) ± 0.003 °C
Lead-wire cable type Enameled copper wire
Lead-wire length 2.4 meters (8 ft)
Lead-wire temperature range 160 °C
Calibration Calibration not included. NVLAP accredited calibration available. See ordering information. |
Calibration Included |
None |
Data Logging |
No |
Interfaces I/O |
Other |
Product Weight |
5.95 LBS |
Product Height |
7.87 IN |
Product Length |
24.02 IN |
Product Width |
12.2 IN |
Shipping Weight |
6.55 LBS |
ECCN Number |
3A992 |
Country of Origin |
China |
Shipping Height |
9.84 IN |
Shipping Length |
24.02 IN |
Shipping Width |
12.2 IN |
UPC |
095969545174 |
|
|
|
|