Fluke Calibration 5611T-G
Catalog #: 2740045
Teflon Reference, Bare Thermistor Probe, Gold Pins
- Overview
- Specifications
- Summary
- Cases
Product Highlights
- Gold pin connector tips
- Probe tip: 0.12 inch diameter and 1.1 inch probe tip length
- Short-term accuracy to +/- 0.01 °C; one-year drift < +/-0.01 °C
- Accredited NVLAP calibration optional
- PTFE encapsulated thermistor
Each Unit Includes
- Coefficient Sheet
- Certification Sheets
Specifications |
Parameter |
Value |
Maximum temperature |
100 °C |
Minimum temperature |
0 °C |
Nominal resistance 25 °C |
10,000 W ± 0.01 |
Temperature coefficient |
Negative (NTC) |
Length of sensor |
13 mm (0.51 in) |
Length of sheath |
2.79 cm (1.10 in) |
Diameter of sheath |
3.0 mm (0.12 in) |
Sheath material |
Teflon |
Immersible length of instrument |
~ 6 m (20 ft) |
Short term stability maximum temp (no less than three thermal cycles) |
<6 mK |
Long term stability maximum temp (no less than 100 hrs @ max temp) |
<6 mK |
Hysteresis (if different from short term stability) |
<1 mK |
Self heating (DC in 25°C bath) |
4 mW/°C typical |
Self heating (DC in room temperature air) |
.014 mW/°C typical |
Time constant (room temperature to 70°C bath) |
4 seconds typical |
Time constant (70°C bath to room temperature) |
30 seconds typical |
Susceptibility to vibration (lo med hi) |
Low |
Lead wire cable type |
4 wire, 30 AWG, 25/44, clear FEP Teflon insulation |
Lead wire outside diameter |
1.75 mm (0.069 in) |
Lead wire length |
6.1 m (20 ft) |
Standard calibration |
NIST traceable calibration, R vs. T table with 0.1 °C increments, interpolation equation furnished |
Calibration uncertainty |
Table and equation are accurate to ± 0.01 °C |
Optional calibration |
Certificate of NVLAP-accredited calibration, Data supplied at 6 points over 100 °C span, complies with NCSL/ISO/IEC 17025:2000 and ANSI/ NCSL Z540-1-1994 |
Termination |
Specify when ordering |
Country of Origin |
United States |
|
|
|
|