|
|
Fluke Calibration 5627A-9-B
Catalog #: 3003720
Precision Industrial PRT Probe, 9 X 3/16 INCH OD
- Overview
- Specifications
- Summary
- Cases
Product Highlights
- 5627A-9-B (bare wire)
- 229 mm x 4.7 mm (9 x 3/16 in)
- Temperature Range –200 °C to 300 °C
- Vibration and shock resistant
- NVLAP-accredited calibration included, lab code 200706-0
Each Unit Includes
- Coefficient Sheet
- User Manual CD
- Certification Sheets
Specifications |
Resistance |
Nominal 100 Ω |
Temperature Coefficient |
0.00385 Ω/Ω/ °C nominal |
Temperature Range |
-200 °C to 420 °C ( and to 300 °C; transition and cable temperature: 0 °C to 150 °C) |
Drift Rate (k=2) |
± 0.04 °C at 0 °C after 100 hours at 420 °C |
Sheath Material |
316 Stainless Steel |
Leads |
PTFE-insulated, nickel-plated stranded copper, 22 AWG |
Termination |
Specify. See Ordering Information. |
Time Constant |
Four seconds maximum for 63.2 % response to step change in water moving at 3 fps |
Bending Radius |
Sheath may be ordered with a bend on a minimum radius of 19 mm (3/4 in) except for 50 mm (2 in) area of sheath near tip. (Fluke lab requires 20 cm [8 in] of unbent sheath to re-calibrate.) |
Calibration |
Includes manufacturer's NVLAP-accredited (lab code 200706-0) calibration and table with R vs. T values in 1 °C increments from -196 °C to 500 °C (to 300 °C for and ). ITS-90 coefficients included |
Immersion |
At least 100 mm (4 in) recommended |
Calibrated Accuracy† (k=2) |
± 0.026 °C at -196 °C
± 0.046 °C at 0 °C
± 0.077 °C at 200 °C
± 0.124 °C at 420 °C |
Size |
: 305 mm x 6.35 mm (12 in x 1/4 in)
: 229 mm x 4.7 mm (9 in x 3/16 in)
: 152 mm x 4.7 mm (6 in x 3/16 in) |
†Includes calibration uncertainty and 100 hr drift.
|
Accessories/Probes/Temperature Probes Template |
Application/Properties |
Piercing |
Max Temperature |
300 C |
Min Temperature |
-200 C (-328 F) |
Wire Termination |
Bare Wire |
Insulation Material |
Stainless Steel |
Probe Length |
9 IN |
Unique Features |
Resistance Nominal 100 Ω
Temperature Coefficient 0.00385 Ω/Ω/ °C nominal
Temperature Range −200 °C to 420 °C (5627A-6 and 5627A-9 to 300 °C; transition and cable temperature: 0 °C to 150 °C)
Drift Rate (k=2) ± 0.04 °C at 0 °C after 100 hours at 420 °C
Sheath Material 316 Stainless Steel
Leads PTFE-insulated, nickel-plated stranded copper, 22 AWG
Termination Specify. See Ordering Information.
Time Constant Four seconds maximum for 63.2 % response to step change in water moving at 3 fps.
Bending Radius
Sheath may be ordered with a bend on a minimum radius of 19 mm (3/4 in) except for 50 mm (2 in) area of sheath near tip. (Fluke lab requires 20 cm [8 in] of unbent sheath to re-calibrate.)
Calibration
Includes manufacturer's NVLAP-accredited (lab code 200706-0) calibration and table with R vs. T values in 1 °C increments from -196 °C to 500 °C (to 300 °C for 5627A-6 and 5627A-9).
ITS-90 coefficients included.
Immersion
At least 100 mm (4 in) recommended
Calibrated Accuracy† (k=2 |
Warranty |
1 YEARS |
Calibration Included |
Factory Calibration |
Data Logging |
No |
Interfaces I/O |
Other |
Product Weight |
7.94 LBS |
Product Height |
12.01 IN |
Product Length |
24.02 IN |
Product Width |
12.01 IN |
Shipping Weight |
21.91 LBS |
ECCN Number |
3A992 |
Country of Origin |
United States |
Shipping Height |
3.94 IN |
Shipping Length |
24.87 IN |
Shipping Width |
12.99 IN |
UPC |
95969403306 |
|
|
|
|