Fluke 80T-150UA
Catalog #: 3190456
Universal Temperature Probe
Product Highlights
- Compatible with Fluke DMM
- High accuracy, fast reading for low voltage (below 24 V AC, 60 V DC) applications
- Measurement range: -50 to 150°C
- Output: 1 mV/°C or 1 mV/°F (switchable)
- One year warranty
- Country of Origin Taiwan
General |
Weight |
150 g (5.3 oz) |
Overall Length |
53.8 inch (1.36 m) |
Battery |
Standard 9 V battery (NEDA #1604,6F22,006P) |
Battery Life |
600+ hours, typical (Alkaline Battery), 6.5 V minimum |
Output Termination |
Standard 0.75-inch spaced double banana plug |
Probe Material |
Glass-filled valox |
Probe Size |
0.6 inch maximum diameter |
Tip Material |
Nickle Plated Copper |
Tip Size |
0.31 cm (0.12 inch diameter) |
Electrical |
Measurement Range |
-50 to +150 °C
-58 to +302 °F |
Sensitivity (80T-150UA output) |
1 mV dc / °C or °F |
Voltage Standoff |
60 V dc or 30 V ac rms (42.4 V peak) |
Settling Time |
9.5 seconds to settle within 2 ° for a 50 ° change |
RF Field |
The 80T-150UA product is unspecified in environments with EMC fields greater than 100 mV/meter. ESD performance is specified as Criteria C per EN61326. |
Environmental |
Ambient Operating Range for Temperature Module |
0 to +50 °C (+32 to +122 °F) |
Maximum Temperature Probe Body and Cable |
+70 °C (+160 °F) See Probe Limitations |
Maximum Temperature Probe Body and Cable |
+70 °C (+160 °F) See Probe Limitations |
Minimum Temperature Probe Body and Cable |
Minimum Operating Temperature for Cable |
-12 °C (+10 °F) |
Minimum Operating Temperature for Probe Body |
-40 °C (-40 °F) See Probe Limitations |
Storage Temperature for Unit |
-40 to +70 °C (-40 to +160 °F) |
Humidity |
0 % to 90 % (0 °C to +35 °C)
0 % to 70 % (+35 °C to +50 °C) (noncondensing) |
Altitude |
Operating: =3048 m ( =10,000 ft) |
Storage |
≤15240 m (=50,000 ft) |
Application Force |
15 pounds maximum (probe tip to measured surface) |
Accuracy (Applies for one year after purchase or calibration) |
AMBIENT °C |
ACCURACY |
+15 to +35 °C |
±1 °C from 0 to +100 °C, decreasing linearly to ±3 °C at -50 and +150 °C |
0 to +15 °C and +35 to +50 °C |
±2 °C from 0 to +100 °C, decreasing linearly to ±4 °C at -50 and +150 °C |
AMBIENT °C |
ACCURACY |
+59 to +95 °F |
±1.8 °F from +32 to +212 °F, decreasing linearly to ±5.4 °F at -58 and +302 °F |
+32 to +59 °F and +95 to +122 °F |
±3.6 °F from +32 to +212 °F, decreasing linearly to ±7.2 °F at -58 and +302 °F |
|
|
|
|