Fluke I310S
Catalog #: 2840309
Current Probe
Product Highlights
- Current Ranges: 30 A and 300 A ACRMS or ±45 A and 450 A DC
- Measuring Ranges: ±45 A and 450 A dc
- Output Sensitivity: 10 mV/A (40 A) to 1 mV/A (400 A)
- Resolution: ±50 mA (40 A) to ±100 mA (400 A)
- Load Impedance: > 10 k Ohms and £ 100 pF
- Frequency Range (small signal): DC to 20 kHz (- 3 dB)
- Phase Shift below 1 kHz, < 2 degrees
- Temperature Coefficient: ±0.1% of reading / °C
- Power Supply: 9 V Alkaline
- Country of Origin China
| Electrical Characteristics |
| Current Ranges |
30 A and 300 A ACRMS or ±45 A and 450 A DC |
| ACRMS or DC |
±45 A and 450 A |
| Inrush Current |
600A ACRMS MAX |
| Output Sensitivity |
10 mV/A (30 A)
1 mV/A (300 A) |
| Accuracy (at +23°C) |
(30A range) ±1% of reading ±50 mA
(300A range) ±1% of reading ±300 mA |
| Bandwidth to meet accuracy specification |
1 kHz |
| Phase Shift below 1 kHz |
< 2 degrees |
| Resolution |
±50 mA (30 A)
±100 mA (300 A) |
| Load Impedance |
> 10 k Ohms and =100 pF |
| Conductor Position Sensitivity |
±1.5% relative to centre reading |
| Frequency Range |
(small signal) DC to 20 kHz (- 3 dB) |
| Temperature Coefficient |
±0.1% of reading / °C |
| Power Supply |
9 V Alkaline, MN1604/PP3 |
| Battery Life |
30 Hours, low battery indicator |
| Working Voltage (see Safety Standards section) |
300 V ACRMS or DC |
| General Characteristics |
| Maximum Conductor Size |
19 mm diameter |
| Output Connection |
Safety BNC connector
Supplied with safety 4mm adaptor |
| Output Zero |
Manual adjust via thumbwheel |
| Cable Length |
2 metres |
| Operating Temperature Range |
10 to +50 °C |
| Storage Temperature Range (with battery removed) |
-20 to +85 °C |
| Operating Humidity |
15% to 85% (non-condensing) |
| Weight |
250 g |
| Safety Standards |
| BS EN 61010-1 |
2001 |
| BS EN 61010-2-032 |
2002 |
| BS EN 61010-031 |
2002 |
| 300 VRMS, Category III, Pollution Degree 2 |
|
| Note |
Use of the probe on un-insulated conductors is limited to 300 V ACRMS or DC and frequencies below 1 kHz. |
| EMC Standards |
BS EN 61236 1998 +A1, A2, & A3 |
|
|
|
|