Fluke Calibration 1523-156
Catalog #: 3376516
Handheld Thermometer Readout, 1 Channel
- Overview
- Specifications
- Summary
- Cases
- Software
- Temp. Probes
- Accessories
Product Highlights
- Three sensor types
- PRTs: –200 °C to 1000 °C
- Thermocouples –200 °C to 2315 °C
- Precision thermistors: –50 °C to 150 °C
- Accuracy
- PRTs: up to ±0.011 °C
- Thermocouples : ±0.24 °C for J,K,L,M
- Precision thermistors: ±0.002 °C
- Fast mode
- PRTs: 0.45 seconds per sample
- Thermocouples: 0.3 second per sample
- Precision thermistors: 0.3 seconds per sample
- Graphical display
- 128x64 backlit LCD graphic display
- Plot and scale trends in real time
- Simultaneous dual channel readings
Each Unit Includes
- NIST Traceable Certificate of Calibration
- Users Guide
- CD-ROM (contains technical manual)
- 12 V dc Universal Power Supply
- RS-232 Cable
- 9940 I/O ToolKit Software
Specifications |
|
1523 |
1524 |
Input Channels |
1 |
2 |
Logging |
25 readings with statistics |
25 readings with statistics
15,000 time and date stamped |
Sample interval (normal): |
1 second |
1 second (simultaneous measurement) |
Typical sample interval (fast mode)*: |
0.3 seconds |
0.3 seconds |
Sensor types |
PRTs, RTDs, Thermistors, and Thermocouples |
Thermocouple types |
C,E,J,K,L,M,N,T,U,B,R,S |
Operating temperature |
–10 °C to 60 °C (Best accuracy 13 °C to 33 °C) |
Power requirements |
3 AA alkaline batteries |
Safety |
EN 61010-1:2001, CAN/CSA C22.2 No. 61010.1-04 |
Size (HxWxL) |
96 mm x 200 mm x 47 mm
(3.75 in x 7.9 in x 1.86 in) |
Weight |
0.65 kg (1.4 lb) |
Environmental conditions for best accuracy: 13 °C to 33 °C |
Millivolt range and accuracy |
–10 mV to 75 mV
± (0.005 % + 5 V) |
Resistance range and accuracy |
0 Ω to 400 Ω ± (0.004 % + 0.002 Ω)
200 Ω to 50 kΩ ± (0.01 % + 0.5 Ω)
50 kΩ to 500 kΩ ± (0.03 %) |
Temperature coefficient, voltage ( –10 °C to 13 °C, + 33 °C to 60 °C) |
Temperature coefficient
( –10 °C to 13 °C , +33 °C to 60 °C):
± (0.001 %/°C + 1 V/°C) |
Temperature Coefficient, Resistance ( –10 °C to 13 °C, + 33 °C to 60 °C) |
0.0008 %/ °C + 0.0004 Ω (0 Ω to 400 Ω)
0.002 %/ °C + 0.1 Ω (0 Ω to 50 kΩ)
0.06 %/ °C + 0.1 Ω (50 kΩ to 500 kΩ) |
Excitation current, resistance |
1 mA (0 Ω to 400 Ω)
10 µA (0 Ω to 50 kΩ)
2 µA (50 kΩ to 500 kΩ) |
*See technical manual for sample interval details by sensor type and number of inputs. |
1523/24 Accuracies with selected probes (± °C) |
T(°C) |
5616-12 |
5615-6 |
5627A-12 |
5610-9 |
-200 |
0.014 |
0.025 |
0.027 |
n/a |
0 |
0.021 |
0.021 |
0.049 |
0.009 |
100 |
0.027 |
0.028 |
0.065 |
0.030 |
300 |
0.040 |
0.043 |
0.103 |
n/a |
420 |
0.050 |
n/a |
0.130 |
n/a |
Includes readout accuracy, probe calibration, and probe drift |
|
Temperature, thermocouples |
Type |
Range |
Measure accuracies |
B |
600 °C to 800 °C
800 °C to 1000 °C
1000 °C to 1800 °C |
0.85 °C
0.68 °C
0.57 °C |
C |
100 °C to 550 °C
550 °C to 2300 °C |
0.32 °C
0.71 °C |
E |
–200 °C to 0 °C
0 °C to 950 °C |
0.52 °C
0.22 °C |
J |
–200 °C to 0 °C
0 °C to 1200 °C |
0.52 °C
0.23 °C |
K |
–200 °C to 0 °C
0 °C to 1370 °C |
0.61 °C
0.24 °C |
L |
–200 °C to 0 °C
0 °C to 900 °C |
0.36 °C
0.23 °C |
M |
–20 °C to 0 °C
0 °C to 400 °C
400 °C to 1400 °C |
0.26 °C
0.25 °C
0.22 °C |
N |
–200 °C to 0 °C
0 °C to 1300 °C |
0.72 °C
0.28 °C |
R |
–20 °C to 0 °C
0 °C to 500 °C
500 °C to 1750 °C |
1.09 °C
0.97 °C
0.49 °C |
S |
–20 °C to 0 °C
0 °C to 500 °C
500 °C to 1750 °C |
1.05 °C
0.95 °C
0.56 °C |
T |
–200 °C to 0 °C
0 °C to 400 °C |
0.60 °C
0.25 °C |
U |
–200 °C to 0 °C
0 °C to 400 °C |
0.54 °C
0.24 °C |
Resolution: |
B, C, E, J, K, L, M, N, R, S, T, U: |
0.10 °C |
Note 1: Accuracies are based on internal reference junction compensation. Refer to technical manual for equivalent accuracies with an external reference junction. |
|
Calibration Included |
NIST Calibration |
Safety Approval |
CSA |
Data Logging |
Yes |
Interfaces I/O |
RS-232 |
Product Weight |
1.4 LBS |
Product Height |
47 MM (1.85 IN) |
Product Length |
96 MM (3.78 IN) |
Product Width |
200 MM (7.87 IN) |
ECCN Number |
EAR99 |
HTS/Schedule B Number |
9025.19.8080 |
Country of Origin |
United States |
|
Cases
Software
Temp. Probes
Accessories
|
|
|