Fluke Calibration 5618B-12-A
Catalog #: 4623267
Secondary PRT Probe, 12 X 1/8 INCH OD
- Overview
- Specifications
- Summary
- Cases
Product Highlights
- 5618B-12-A Probe, Secondary PRT, 6X1/8 Inch
- Small diameter sheath, 3.2mm (0.125 in)
- Excellent stability
- Includes ITS-90 coefficients
- NVLAP-accredited calibrated from –200°C to 500°C, lab code 200706-0
Each Unit Includes
- Coefficient Sheet
- User Manual CD
- Certification Sheets
- INFO-CON Addendum
Specifications |
Resistance |
Nominal 100 Ω at 0 °C |
Temperature
Coefficient |
0.003923 Ω/Ω/ °C nominal |
Temperature
Range |
–200 °C to 500 °C
(–200 °C to 300 °C for 5618B-6-X) |
Drift Rate |
± 0.1 °C when used periodically to 500 °C |
Sheath Material |
316 SST |
Leads |
22 AWG PTFE, 6’ |
Termination |
Specify |
Hysteresis |
Less than 0.01 °C at 0 °C when using –196 °C and
420 °C as the end points. |
Time Constant
|
Four seconds maximum for 63.2 % response to
step change in water moving
at 3 fps. |
Thermal EMF |
Less than 25 mV at 420 °C |
Calibration
|
Includes manufacturer’s NVLAP-accredited
calibration w/ITS-90 coefficients, R vs. T values in
1 °C increments, lab code 200706-0 |
Size
|
5618B-12: 305 mm L x 3.2 mm dia. (12 x 1/8 in)
5618B-9: 229 mm L x 3.2 mm dia. (9 x 1/8 in)
5618B-6: 152 mm L x 3.2 mm dia. (6 x 1/8 in) |
Calibrated
Accuracy† (k=2) |
± 0.026 °C at –200 °C
± 0.066 °C at 0 °C
± 0.219 °C at 500 °C |
†Includes
calibration
uncertainty and
drift specification. |
NVLAP† Calibration Uncertainty |
Temperature |
Expanded Uncertainty (k=2) |
–196 °C |
0.024 °C |
–38 °C |
0.011 °C |
0 °C |
0.010 °C |
200 °C |
0.018 °C |
420 °C‡ |
0.029 °C |
Note: Calibration uncertainties depend on the uncertainties of the lab performing the calibration. Subsequent calibrations of this probe performed with different processes, at different facilities, or with changed uncertainty statements may state different uncertainties. †Lab code 200706-0 ‡300 °C for 5618B-6 |
|
Country of Origin |
United States |
|
|
|
|